We report here the design of a novel immunosensor and its application for celiac disease diagnosis, based on an electrogenerated chemiluminescence (ECL) readout, using membrane-templated gold nanoelectrode ensembles (NEEs) as a detection platform. An original sensing strategy is presented by segregating spatially the initial electrochemical reaction and the location of the immobilized biomolecules where ECL is finally emitted. The recognition scaffold is the following: tissue transglutaminase (tTG) is immobilized as a capturing agent on the polycarbonate (PC) surface of the track-etched templating membrane. It captures the target tissue transglutaminase antibody (anti-tTG), and finally allows the immobilization of a streptavidin-modified ruthenium-based ECL label via reaction with a suitable biotinylated secondary antibody. The application of an oxidizing potential in a tri-n-propylamine (TPrA) solution generates an intense and sharp ECL signal, suitable for analytical purposes. Voltammetric and ECL analyses evidenced that the ruthenium complex is not oxidized directly at the surface of the nanoelectrodes; instead ECL is generated following the TPrA oxidation, which produces the TPrA•+ and TPrA• radicals. With NEEs operating under total overlap diffusion conditions, high local fluxes of these reactive radicals are produced by the nanoelectrodes in the immediate vicinity of the ECL labels, so that they efficiently generate the ECL signal. The radicals can diffuse over short distances and react with the Ru(bpy)32+ label. In addition, the ECL emission is obtained by applying a potential of 0.88 V versus Ag/AgCl, which is about 0.3 V lower than when ECL is initiated by the electrochemical oxidation of Ru(bpy)32+. The immunosensor provides ECL signals which scale with anti-tTG concentration with a linearity range between 1.5 ng·mL-1 and 10 μg·mL-1 and a detection limit of 0.5 ng·mL-1. The sensor is finally applied to the analysis of anti-tTG in human serum samples, showing to be suitable to discriminate between healthy and celiac patients.

We report here the design of a novel immunosensor and its application for celiac disease diagnosis, based on an electrogenerated chemiluminescence (ECL) readout, using membrane-templated gold nanoelectrode ensembles (NEEs) as a detection platform. An original sensing strategy is presented by segregating spatially the initial electrochemical reaction and the location of the immobilized biomolecules where ECL is finally emitted. The recognition scaffold is the following: tissue transglutaminase (tTG) is immobilized as a capturing agent on the polycarbonate (PC) surface of the track-etched templating membrane. It captures the target tissue transglutaminase antibody (anti-tTG), and finally allows the immobilization of a streptavidin-modified ruthenium-based ECL label via reaction with a suitable biotinylated secondary antibody. The application of an oxidizing potential in a tri-n-propylamine (TPrA) solution generates an intense and sharp ECL signal, suitable for analytical purposes. Voltammetric and ECL analyses evidenced that the ruthenium complex is not oxidized directly at the surface of the nanoelectrodes; instead ECL is generated following the TPrA oxidation, which produces the TPrA•+ and TPrA• radicals. With NEEs operating under total overlap diffusion conditions, high local fluxes of these reactive radicals are produced by the nanoelectrodes in the immediate vicinity of the ECL labels, so that they efficiently generate the ECL signal. The radicals can diffuse over short distances and react with the Ru(bpy)32+ label. In addition, the ECL emission is obtained by applying a potential of 0.88 V versus Ag/AgCl, which is about 0.3 V lower than when ECL is initiated by the electrochemical oxidation of Ru(bpy)32+. The immunosensor provides ECL signals which scale with anti-tTG concentration with a linearity range between 1.5 ng·mL–1 and 10 μg·mL–1 and a detection limit of 0.5 ng·mL–1. The sensor is finally applied to the analysis of anti-tTG in human serum samples, showing to be suitable to discriminate between healthy and celiac patients.

A Sensitive Electrochemiluminescence Immunosensor for Celiac Disease Diagnosis Based on Nanoelectrode Ensembles

HABTAMU, HENOK BAYE;SILVESTRINI, MORENA;UGO, Paolo
2015-01-01

Abstract

We report here the design of a novel immunosensor and its application for celiac disease diagnosis, based on an electrogenerated chemiluminescence (ECL) readout, using membrane-templated gold nanoelectrode ensembles (NEEs) as a detection platform. An original sensing strategy is presented by segregating spatially the initial electrochemical reaction and the location of the immobilized biomolecules where ECL is finally emitted. The recognition scaffold is the following: tissue transglutaminase (tTG) is immobilized as a capturing agent on the polycarbonate (PC) surface of the track-etched templating membrane. It captures the target tissue transglutaminase antibody (anti-tTG), and finally allows the immobilization of a streptavidin-modified ruthenium-based ECL label via reaction with a suitable biotinylated secondary antibody. The application of an oxidizing potential in a tri-n-propylamine (TPrA) solution generates an intense and sharp ECL signal, suitable for analytical purposes. Voltammetric and ECL analyses evidenced that the ruthenium complex is not oxidized directly at the surface of the nanoelectrodes; instead ECL is generated following the TPrA oxidation, which produces the TPrA•+ and TPrA• radicals. With NEEs operating under total overlap diffusion conditions, high local fluxes of these reactive radicals are produced by the nanoelectrodes in the immediate vicinity of the ECL labels, so that they efficiently generate the ECL signal. The radicals can diffuse over short distances and react with the Ru(bpy)32+ label. In addition, the ECL emission is obtained by applying a potential of 0.88 V versus Ag/AgCl, which is about 0.3 V lower than when ECL is initiated by the electrochemical oxidation of Ru(bpy)32+. The immunosensor provides ECL signals which scale with anti-tTG concentration with a linearity range between 1.5 ng·mL–1 and 10 μg·mL–1 and a detection limit of 0.5 ng·mL–1. The sensor is finally applied to the analysis of anti-tTG in human serum samples, showing to be suitable to discriminate between healthy and celiac patients.
2015
87
File in questo prodotto:
File Dimensione Formato  
2015-ECL-NEE anti-tTG-Anal Chem-Habtamu.pdf

non disponibili

Descrizione: Articolo
Tipologia: Versione dell'editore
Licenza: Accesso chiuso-personale
Dimensione 2.39 MB
Formato Adobe PDF
2.39 MB Adobe PDF   Visualizza/Apri

I documenti in ARCA sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10278/3664733
Citazioni
  • ???jsp.display-item.citation.pmc??? 6
  • Scopus 60
  • ???jsp.display-item.citation.isi??? 53
social impact